Hamilton, 10069701, OxyFerm FDA MS 160

The OxyFerm FDA Memosens is an electrochemical oxygen sensor specifically designed for hygienic applications with stringent requirements, such as those found in the pharmaceutical, biotechnology, and food and beverage industries.
Catalog Number: 10069701
Pricing

Send a RFQ.

View RFQ List
Manufacturer's Pagewww.hamiltoncompany.com/sensors/dissolved-oxygen-sensors/oxyferm-fda/oxyferm-fda-ms?part-number=10069701
Description
The OxyFerm FDA Memosens is a polarographic dissolved oxygen (DO) sensor designed for hygienic applications in the pharmaceutical, biotech, and food & beverage industries. Available in 12 mm shaft diameters, the sensor features an FDA-approved membrane and delivers reliable nanoamp (nA) current output. Built to withstand steam sterilization, autoclaving, and CIP procedures, the OxyFerm FDA Memosens ensures consistent DO measurements in demanding environments. Memosens technology allows upgrading an existing installation with proven Hamilton polarographic DO sensor technology.
Specifications

General

Process ConnectionPG13,5
Sensor FamilyOxyFerm FDA MS
Wetted PartsStainless Steel 1.4435; EPDM (Ethylene propylene elastomer); FKM (Fluorocarbon Elastomer); VMQ (Silicone elastomer); See compliance details in Material Specification document
Electrical ConnectorMemosens
Measuring Range10 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂)
ElectrolyteOxylyte
Drift at Room Temperature< 1% per week
Membrane CapFDA
Oxygen ConsumptionCa. 20 ng/h in air at 25 °C
Temperature SensorNTC 22 kOhm
Current in Air at 25 °C40 - 80 nA
Surface Quality of SteelRa < 0.4 µm (N5)
Electrode systemSilver platinum combination
Polarization Voltage and Time-670 ± 50 mV, ≥ 2 hours
Serial NumberYes
Response Timet98% < 60 s at 25 °C
CertificateMaterial and Performance Test Certificate
ATEX ApprovalII 1 G Ex ia IIC T3/T4/T6 Ga
IECEx ApprovalII 1 G Ex ia IIC T3/T4/T6 Ga
Diameter12 mm
AutoclavableYes
Steam SterilizableYes
CIPYes
Operating Temperature Range0 to 130 °C
Pressure Range bar g0 to 4 bar
Max. CO₂ Partial Pressure0.01 bar
Required Flow≥ 0.03 m/s