The PRP-h5 derives its hydrophobic functionality from a pentafluorinated polymer bead, delivering a selectivity difference from standard silica C18 stationary phases. This gives chromatographers the desired retention characteristics necessary for certain sample types. This difference is especially pronounced for halogenated solutes. The large pore size of 300 Å enables the analysis of protein digests, tryptic digests, and proteomics. Application Examples: Organic compounds: macromolecules (> 2,000 mw), pharmaceuticals, protein digests, tryptic digests, proteomics
Specifications

General

Separation ModeReversed-Phase
Stationary PhasePRP-h5
FormatHPLC Column
Hardware Inner Diameter10 mm
Hardware Length150 mm
Particle Size5 µm
Packing Material TypePSDVB
Pore Size300 Å
Particle Surface Area (m2/g)277
Pore Volume (cm3/g)0.526
Hardware MaterialStainless Steel
Max Pressure5,000 psi
Mobile Phase LimitspH 1-13, 0-100% aqueous, organic modifier
Buffer Strength0.0-0.5N
Temp Limits5-85°C
RestorationRun a gradient of 100% water to 100% acetonitrile. Repeat three times
Compound ClassificationPeptides/Proteins
UOM1 EACH